Prismaflex TPE 1000 and TPE 2000 Sets In Vitro Performances

General Data Prismaflex TPE 1000 set

Prismaflex TPE 2000 set

Weight

680 g

760 g

Blood volume in set ± 10 %

71 ml

125 ml

Blood volume in plasmafilter

23 ml ±10%

41 ml ±10%

9 kg

adult

Blood pressure drop ±30% (Bovine blood, Htc***32%, Protein concentration 60 g/l, 37°C) Prismaflex TPE 1000 set

Prismaflex TPE 2000 set

QB** = 50 ml/min

14 mmHg - 1.9 kPa



QB** = 100 ml/min

29 mmHg - 3.9 kPa

25 mmHg - 3.3 kPa

Materials

QB** = 180 ml/min

52 mmHg - 6.9 kPa



Plasmafilter hollow fiber: Polypropylene

QB** = 200 ml/min



51 mmHg - 6.8 kPa

Filter housing and headers: Polycarbonate

QB** = 250 ml/min



64 mmHg - 8.5 kPa

Filter potting compound: Polyurethane

QB** = 400 ml/min



101 mmHg - 13.5 kPa

Tubing material: Plasticized polyvinyl chloride (PVC) Cartridge: PETG

*Arterial transmembrane pressure **Access blood flow rate ***Hematocrit

Sterilization mode: EtO (ethylene oxide)

Sieving coefficient

Minimal patient weight

Filter operating specifications

(In vivo data, 19 treatments) Albumin = 0.97, IgG = 1, Apolipoprotein B = 0.95, IgM = 0.92

Maximum blood flow rate: Prismaflex TPE 1000 set: 180 ml/min Prismaflex TPE 2000 set: 400 ml/min

Plasma filtration with blood In vitro, values ±20% (Bovine blood at 37°C, Htc 32%, Protein concentration 60 g/l) Plasma filtration is controlled by the Prismaflex system.

Minimum blood flow rate: Prismaflex TPE 1000 set: 50 ml/min Prismaflex TPE 2000 set: 100 ml/min Maximum TMPa* (mmHg/kPa):

Prismaflex TPE 1000 set Prismaflex TPE 2000 set

QB** = 50 ml/min

100 mmHg - 13.3 kPa



QB** = 100 ml/min

140 mmHg - 18.7 kPa

120 mmHg - 16 kPa

QB** = 180 ml/min

190 mmHg - 25.3 kPa



QB** = 200 ml/min



171 mmHg - 22.8 kPa

QB** = 250 ml/min



193 mmHg - 25.7 kPa

QB** = 400 ml/min



246 mmHg - 32.8 kPa

Filter data Nominal physical characteristics Effective surface area: Prismaflex TPE 1000 set: 0.15 m² Prismaflex TPE 2000 set: 0.35 m²

QB = 180 ml/min

40

Fiber wall thickness:

150 μm

30

QB = 100 ml/min

20

QF max = 16 ml/min

0 0

20

40

60

80

100

120

TPE 1000-2000 2016_Nordic.indd 2

160

180

160

107143 115314 107144 115315

N° units/box 4 4

QF max = 147 ml/min

QB = 400 ml/min

120 100

QF max = 114 ml/min

QB = 300 ml/min

80

QB = 200 ml/min

60

QF max = 79 ml/min

QF 42 = 114 ml/min

40

QB = 100 ml/min

20 0 0

50

100

150

TMP (mmHg)

www.baxter.se

140

Prismaflex TPE 2000 set

Plasma filtration rate (ml/min)

Code N°

Box 63 164 94 Kista T: 08-632 64 00

QF max = 29 ml/min

QB = 50 ml/min

10

140

Ordering Information

Sweden: Baxter Medical AB

QF max = 49 ml/min

Denmark: Baxter A/S Tobaksvejen 2A 2860 Søborg T: 4816 6400 www.baxter.dk

Finland: Baxter Oy PL 119 00181 Helsinki T: (09) 8621 111 www.baxter.fi

Norway: Baxter AS Gjerdrumsvei 11 0484 Oslo T: 22 58 48 00 www.baxter.no

200

250

SE/MG120/16-0002, NO/MG120/16-0003, FI/MG120/16-0002, DK/MG120/16-0002

330 μm

Prismaflex TPE 2000 set Prismaflex TPE 2000 set CKT

50

TMP (mmHg)

Fiber internal diameter (wet):

Prismaflex TPE 1000 set Prismaflex TPE 1000 set CKT

60

Plasma filtration rate (ml/min)

Prismaflex TPE 1000 set